Evaluation of ABX Pentra DX 120 and Sysmex XE-2100 in umbilical cord blood

By H. KIM, M. HUR, S.-G. CHOI, B.-H. KIM, G, H. S. KWON, I. S. SOHNH.-W. MOON, Y.-M. YUN, H. S. HWAN-
08/07/2020

250 0
categin thumb1

Cord blood (CB) is an important source of hematopoietic stem cells and reflects the hematologic status of neonates. ABX Pentra DX 120 (Horiba Medical, Montpellier, France) and Sysmex XE-2100 (Sysmex, Kobe, Japan) were compared in 200 CB specimens. Most of the parameters showed acceptable correlation between the two analyzers. Although WBC differential of both analyzers showed acceptable correlation with manual counts, mononuclear cells (MNC) by ABX Pentra DX 120 better correlated with manual count than MNC by Sysmex XE-2100. NRBC by Sysmex XE-2100 better correlated with manual count than NRBC by ABX Pentra DX 120. ABX Pentra DX 120 showed better flagging performances. DDX better correlated with CD34+ cells than HPC.