Cell counting of body fluids: comparison between three automated haematology analysers and the manual microscope method

By P. DANISE, M. MACONI, A. ROVETTI, D. AVINO, A. DI PALMA, M. GERARDO PIROFALO, C. ESPOSITO-
13/07/2020

157 1
categin thumb1

Haematological analysis of body fluids (BF) specimens can provide clinicians with valuable diagnostic information because it can indicate one of several serious medical conditions. Although up to now the microscopic counting and the differentiation of WBC in a BF smear have been used as a reference. The introduction of semiautomated and automated methods of analysis has reduced interoperator variability and improved turnaround time and precision. The aim of our study was to evaluate the accuracy and the correlation between the three methods and with the reference method. The introduction of automated method for BF analysis is more and more useful in the routine job of a laboratory analysis. It is therefore very important to evaluate the performance of the different automated haematology technologies, because there is a lack of literature in this field. The comparison between the Pentra DX 120, the other technologies and the manual counting showed instrumental overlapping capabilities.